1m3 Formaldehyde Emission Test Chamber
Guang Dong LI KE XING Spring Machinery Technology Co., Ltd.- Power Supply:220V, 50Hz
- Certification:ISO
- Warranty:1 Year
- Transport Package:Wooden Case
- Specification:w850*D990*H210(Or customized according to customer
- Origin:Made in China
Base Info
- Model NO.:formaldehyde test chamber
- HS Code:8479899990
Description
Project
performance parameter
Box volume(m³)
1m³
Main technical indicators
Studio size
850*1080*1100/550*600*680
Outer case dimension
1460*1225*1572/1100*796*1520
Internal dimensions (about)mm
w850*D990*H210(Or customized according to customer requirements)
Temperature and Humidity Adjustment Range
temperature range:+15ºC~+40ºC
Humidity range:30%~80% RH
fluctuation
Temperature:≤0.5ºC
Humidity:±2%
Uniformity
Temperature:≤1ºC
Temperature and humidity deviation
Temperature:≤0.8ºC
Humidity:≤±2%
Studio wind speed
The wind speed in the center of the box 0.1-1 m/s.(adjustable wind speed)
Ambient Temperature
0ºC~35ºC
Air velocity
0.1m/s~0.3m/s;accuracy≤0.05m/S
Air replacement rate
0.1~1.5time/h
Must be able to maintain(0.1±0.05)time/h/S
Relative barotropic maintenance
10±5pa(online display, recording)
Flow meter accuracy
Glass Rotor Flow meter with accuracy not less than 2.5 grades
closed
When 1000Pa is over-pressurized, the leakage of gas is less than 1 x10-3 m/min or the flow difference between inlet and outlet gas is less than 1%. It is equipped with a micro-pressure sensor, which can monitor the sealing condition of the cabin in real-time according to the pressure.
Stabilization time
Temperature stabilization time is less than 40 minutes, humidity stabilization time is less than 80 minutes, and equipment can run continuously for no less than 30 days.
Background index
Background concentration of formaldehyde:≤0.006mg/m3
Noise
If it is less than or equal to 65 decibels, there should be no resonance sound in the door.
Other
Ensure that no liquid or solid water is present in the tank.
Structural mode
Four corners of the inner tank are designed with large chamfer, and the working cabin is fully welded and welded to ensure sealing performance.
Box body
Features
Material composition
Outer wall material: 1.5MM cold rolled steel sheet electrostatic spraying, the surface is white paint.
Inner wall material: 1.2mm SUS304 mirror stainless steel
Surface finish:III D,All internal interfaces are polished
Insulation material:Natural mineral wool, thickness not less than 80 mm, good insulation
Sealing materials:Fluorine rubber, polytetrafluoroethylene, etc. do not release VOC organic matter
Connecting Pipeline: Intake Part in Contact with Inner Cabin Need or PTFE Material, Cannot Use Organic Material
All materials in the box will not adsorb and release formaldehyde and VOC.
Recirculating fan
Select all-metal blade axial flow fan with oil-free bearing; install two fans at the bottom of the front and rear side (fan plate position down, fan diameter edge distance from the bottom 150 MM), air outlet from the top porous plate, circulating air to facilitate disassembly and cleaning, add large chamfer to the diamond section, do not hurt hands, easy to clean.
Studio
Work cabin integrated forming, seamless welding, weld surface polishing, no dead angle, seamless
Hatch door
Hinged door, multi-point compression type, adjustable clearance (single door)
Observation window
Front door central position, built-in three-layer hollow glass, with heating body, with anti-condensation function;
Power Supply
AC220V380V 50Hz
Sampling hole
Located on the left side of the equipment, there are three air sampling points in the cabin. The sampling port is brass material, which can match the latex pipe connected to 6*9.
Sampler pumping speed(L/min)
0.1~2.5 adjustable can be stabilized at 2.0L/min bracket
Floodlight
The LED lamp belt is designed on the outer cabin for easy observation of the samples in the cabin.
Air intake control
Flow meter needle valve, air intake uniformly released in the working chamber with a porous tube
Intake air purification
Series Air Purification Device (Activated Carbon) behind Air Pump
Cryogen
Environmental protection refrigerant R404a
humidity control
ROTRONIC Temperature and Humidity Sensor (Imported), Temperature 0-70 C, Humidity 0-100% RH
Man-machine interface
Chinese, color LCD (7 inches) display, touch key setting
Indicating accuracy
Temperature 0.1 ºC, Humidity 0.1% RH, Time 1 s
Display item
Temperature, measured temperature, set humidity, measured humidity, dew/point set temperature, dew/point measured humidity, total operation time, including historical data query, real-time curve and historical data playback, fault record
Set item
Working cabin temperature, working cabin humidity, dew/point temperature, working start time
Recording function
Work cabin temperature, humidity, dew/point temperature; provide USB interface, you can use U disk for storage, backup
Data Correction
Temperature, humidity and dew point humidity of working cabin can be corrected
Other
It has the function of setting permission level to avoid user misoperation
security
Working cabin
Overheat alarm protector
Refrigeration compressor
Fan over pressure protector and overload protector
Grounding protection
Safe grounding protection for structural parts, insulation resistance of electrical components and equipment housing not less than 0.5Ω·m
heater
Heater Short Circuit protector
Humidifier
Water shortage protector of humidifier
condenser
Water shortage protector of condenser