Test Instrument

IEC60156 Transformer Oil Breakdown Voltage Bdv Tester Insulating Oil Bdv Tester Gdyj-502

Chongqing Gold Mechanical & Electrical Equipment Co., Ltd.
  • Output Voltage:AC 0V~100kv
  • Rate of Voltage Rise:0.5kv,S~5kv,S±5%
  • Max Test Times:9
  • Stir Time:0~999s
  • Waiting Time:0~999s
  • Test Error:3%Rdg+0.3%Fs

Base Info

  • Model NO.:GDYJ-502
  • Display:256´128 LCD
  • Maximum Storage Capacity:100 Groups of Test Data
  • Power Supply:AC220V±10%, 50Hz,60Hz
  • Transport Package:Plywood Case
  • Specification:55kg
  • Trademark:GOLD
  • Origin:Chongqing, China
  • HS Code:9030849000
  • Production Capacity:200piece,Month

Description

Size

620 mm*430mm*330mm

Weight

41kg

 

Functions:

  • Breakdown test of user-defined;

  • Proof test;

  • IEC60156/IS6792/BS5874 test;

  • ASTM D1816 standard test;

  • ASTM D877 standard test;

  • 5 minutes standard test;

  • Manual test;

  • Calibration interface for customers;

  • Uploading data to PC and making word report automatically;

  • Result print;

 

Features:

  • LCD screen can display test procedure and test result;

  • All the human interaction operations complete by the rotating mouse;

  • It is very easy for customers to measure the breakdown voltage of all kinds of oil;

  • You can print the test result by the micro printer internal the tester;

  • Meanwhile, you can upload the test data to PC from RS232 for data analyzing;

  • The special circuit design makes the tester has a good performance at stability and anti-noise;


    Testing Process

    IEC60156 Transformer Oil Breakdown Voltage Bdv Tester Insulating Oil Bdv Tester Gdyj-502
    IEC60156 Transformer Oil Breakdown Voltage Bdv Tester Insulating Oil Bdv Tester Gdyj-502
    IEC60156 Transformer Oil Breakdown Voltage Bdv Tester Insulating Oil Bdv Tester Gdyj-502