Auto Diagnostic Thickness Test for Non-Woven Fabric
Hangzhou Yante Science and Technology Co., Ltd.- Warranty:1 Year
- Type:Strength Testing Machine
- Load Way:Electronic Load
- Display:Digital
- Control:Computer Control
- Weight:0-100Kg
Base Info
- Model NO.:YT-H4E
- Power Source:AC220V
- Power Supply:AC (100~240)V
- Measuring Range:0-5mm
- Resolution:0.001 mm
- Communication Interface:RS232 Interface
- Print:Built-in Thermal Printer
- Dimension:320(L)*310(W)*400(H)mm
- Net Weight:26.5kg
- Transport Package:Wooden Box
- Trademark:YANTE
- Origin:Hangzhou China
- HS Code:9031809090
Description
Basic Info.
Model NO. YT-H4E Power Source AC220V Power Supply AC (100~240)V Measuring Range 0-5mm Resolution 0.001 mm Communication Interface RS232 Interface Print Built-in Thermal Printer Dimension 320(L)*310(W)*400(H)mm Net Weight 26.5kg Transport Package Wooden Box Trademark YANTE Origin Hangzhou China HS Code 9031809090Product Description
Electronic Thickness Tester is a special instrument for measuring the Non-woven Fabric. It is widely used for measuring the thickness of Non-woven Fabric. It is an ideal testing equipment for papermaking, packaging, scientific research, product quality supervision and inspection industries and departments.
Product Features
One-click test, easy to understand
ARM processor improves the response speed of the instrument, and the calculation is accurate and fast
The rising and falling speed of the probe can be adjusted
Sudden power failure data storage function, data before power failure after power-on is retained and testing can be continued.
Automatic measurement, statistics, and printing of test results
Communication with computer software (purchased separately)
Application Range
Non-woven Fabric
Executive Standard
GB/T 451.3,QB/T 1055,GB/T 24328.2,ISO 534
Technical Parameters
No. | Parameter item | Technical index |
1 | Power Supply | AC(100~240)V,(50/60)Hz 50W |
2 | Working Environment | Temperature(10~35)ºC, Humidity≤ 85% |
3 | Display | 5 inch color touch screen |
4 | Measuring range | (0~10)mm |
5 | Resolution | 0.001mm |
6 | Indication error | ±0.003mm or ±0.5% |
7 | Indication variability | ≤0.003mm or ≤0.5% |
8 | Parallelism between measuring surfaces | ≤0.0004mm |
9 | Contact area | (1000±5) mm² |
10 | Contact pressure | (2±0.1)kPa |
11 | Probe descending speed | (0.5~10)mm/s Adjustable |
12 | Built-in thermal printer | |
13 | Communication Interface | RS232(Default) (USB, WIFI optional) |
14 | Dimension | 325×310×485 mm |
15 | Net Weight | 24kg around |