Capactiance and Tan-Delta Tester FT-10

HV Hipot Electric Co., Ltd.
  • Display:Digital
  • Usage:Electric Component Tester
  • Type:Capacitance Measuring Tester
  • Environment:Room Temperature
  • Warranty:With Warranty
  • Power Source:AC 85V-265V

Base Info

  • Model NO.:FT-10
  • Weight:〉10Kg
  • Installation:Direct Connected
  • Customized:Customized
  • Hv Output:0.5-10kv
  • Capacity:1500va
  • Transport Package:Wooden or Box
  • Specification:ISO9001
  • Trademark:HV Hipot
  • Origin:Wuhan, China
  • Production Capacity:5000,Year

Description

Basic Info.

Model NO. FT-10 Weight 〉10Kg Installation Direct Connected Customized Customized Hv Output 0.5-10kv Capacity 1500va Transport Package Wooden or Box Specification ISO9001 Trademark HV Hipot Origin Wuhan, China Production Capacity 5000/Year

Product Description

General introduction

GD6000A Capacitance and Tan Delta Dissipation Test Set measure the capacitance and dielectric loss factor (tgδ ) of high voltage electric equipment. It can also test dielectric loss of insulating oil test grounded, which is micro controller operated, accurate, small, light weight and reliable.

Features

1. 105*65mm LCD screen.

2. Calendar chip and large storage inside. Can save testing result according to time order, check history record and print the result.

3. with USB port. Check and manage the data and create working report in any PC by our special software.

4. Multiple testing mode, with modes of inside hign voltage, outside high voltage, inner standard, external standard, forward/backward connection, self-excitation. High votage(more than 10kV) dieletric loss test can be made in the situation of external standard outside high voltage.

5. Test full sealed CVT(Capacitive Voltage Transformer) C1 and C2 dieletric loss and capacitance at the same time. Also test CVT transformation ratio and voltage angle difference.

6. High speed sampling signal. Inverter and sampling circuit inside are digitized controled. Output voltage is adjusted continuously.

7. Multiple protection of input voltage fluctuation, output short circuit, over-voltage, over-current, temperature, secure and reliable. Meanwhile, it has the function of grounding testing, that voltage boost is not permitted for non-grounding equipment.

8. No need to dismantle cable to measure dielectric loss and capacitance of CVT.

9. High voltage boost transformer inside, with functions of over zero swich-on and lightning protection. Different level of output high voltage with 0.5kV-10kV, easy wiring.

10. 380V power supply anti-false connection, ensure safety in operation;

11. 45Hz, 50Hz, 55Hz frequency conversion signals, anti-interference.



Technical parameters
 
High Voltage Output0.5 - 10kV
Capacity1500VA
AccuracyTgδ : +-(1.0% of the reading+0.04%),
Cx: +-(1.0% of the reading+5PF)
ResolutionTgδ : 0.01% Cx: 1pF
Measuring range0.1% < tgδ <50%
3PF < Cx <60000PF
10kV, Cx <=30000PF
5kV, Cx <= 60000PF
Power SupplyAC 220V+-10%, 50+-1Hz
Harmonic adaptation<=3%
Conditions of use-15 -50degree Humidity <80%
Dimensions460 (L) *335 (W) *340 (H)
Weight30 kg