Test Instrument

Automatic Aas Analyzer Flame Atomic Absorption Spectrophotometer

Chongqing Gold Mechanical & Electrical Equipment Co., Ltd.
  • Weight:75kg
  • Wavelength Range:190nm~900nm
  • Warranty:12 Months
  • Instrument Power Supply:220V, 3A, 50Hz
  • Detection Accuracy Level:Ppm or Ppb(0.001ppm)
  • Receptivity of The Wavelength:≤0.04nm

Base Info

  • Model NO.:GD-4530F
  • Base Line Stability:≤0.002A,30min (Cu)
  • Characteristic Viscosity:0.02μg,Ml,1% (Cu)
  • Preciseness:0.5%
  • Grating:1800 Lines,mm
  • Transport Package:Standard Wooden Case
  • Specification:700mm*550mm*450mm
  • Trademark:GOLD
  • Origin:Chongqing
  • HS Code:9031809090
  • Production Capacity:Chongqing

Description

0ItemGD-4530F Automatic AAS Analyzer Flame Atomic Absorption Spectrophotometer1Wavelength range190 - 900nm2Spectral bandwidthAutomatically change between 0.1nm,0.2nm,0.4nm,1.0nm and 2.0nm3Accuracy of the wavelength±0.15nm4Receptivity of the wavelength≤0.04nm5Base line stability≤0.002A/30min (Cu)6Characteristic viscosity0.02μg/ml/1% (Cu)7Checking limit0.004μg/ml (Cu)8Preciseness0.5%9Grating1800 lines/mm10InflamerAll-metal titanium burner11AtomizerEffective glass atomizer12Lamp turretsEight lamps13Size and weight700mm×550mm×450mm 75Kg14Power source220V±22V AC 
GD-4530F Automatic AAS Analyzer Flame Atomic Absorption Spectrophotometer
Characteristics: 
 Completely controlled by PC, can flexibly select the flame and graphite furnace atomizer (optional)
 Integrated floated optical platform design can obviously improve the optical system shock resistance and can keep stable though use the optical signal for a long term. 
 Eight light stands can be changed automatically and preheat the eight element lights meantime as well as optimize the working condition of the hollow cathode lamp
 Position adjusting: the best height of the flame burner and can automatically set the front and rear positions
 Fully automated wavelength scanning and peak searching
Complete safety chains protection equipment: the function of warning and automatic safety protection towards the wrong burner, leakage of the gas, under voltage of air and the abnormal flameout
 Deuterium lamp and self-absorption background regulation 
 Data processing: super strong database, possesses more than 500 data self-storage and cut-off storage function, can store the analyzed result with the form of EXCEL and the testing method and the result can be randomly called
 Measuring method: flame absorption method and emission method
 Result printing: parameter printing, data result printing and diagram printing
GD-4530F Automatic AAS Analyzer Flame Atomic Absorption Spectrophotometer
Complete automatic controlling system: 
With the help of the software, the following can be easily achieved
Selection of the element lamp
Up-down-front-rear adjustment of the lifter
 Adjustment of the optical energy
Selection of the slit
 Determination of wavelength scanning and peak searching
 Selection of the atomizer
 Setting of the background deduction method 
 Controlling of the gas flow
 Automatic flaming and flaming out 
Setting of the graphite furnace testing method  
Advanced graphite furnace temperature controlling technology:
 The bringing in of PID technology can effectively overcome the influence on the temperature rising process caused by the voltage fluctuation and the resistance change to make more accurate controlling process. The combination of the 3ms/time fast sampling technique can make more accurate and reliable testing data
 The fast heating capacity can improve the flexibility of the elements further 
Use the ordinary power source of 220V without need of dynamic power of 380V
The maximum procedure heating capacity setting of 20 levels can make a more convenient and easier test of different samples
Three grades adjustable gas flow can accustom to more application needs  
Can timely alarm when the gas and water is stopped and insufficient gas and water, can avoid the equipment damage and measuring error
Safe, reliable and convenient flame system:    
 Epc can control the flow of Acetylene (C2H2) more accurately and it is a kind of system which also can operate easily  
Efficient Atomization system enables a higher sensitivity 
 The whole operation system has a high security as the fire safety system can alarm whenever the electricity is cut off, abnormal flame occurs, a lack of pressure happens or the burner does not match well. And it will automatically turn off the gas, prohibit tempering. Thus it keeps the operation people and equipment from harm and damage. 
Multi-functional software workstation: 
 A workstation that is supported WINDOWS XP/2000/NT
 The rich menu brings great convenience to the use of customer
 Convenient conversion between different menus makes the operation more easily
Various analytical correction methods provide the users more choices 
Basic default parameter settings enable even the beginners can do the normal operation
Flexible storage, editing and printing methods give the user largest support 
Instrument configuration:Atomic Absorption Spectrophotometer   1pc,Air compressor without noise--1pc 
PC workstation: computer, HP inkjet, apparatus operating software (the object can be prevailed on the specific configuration)
A set of accessories (including Acetylene reducing valve, one copper and manganese element lamp and air filter) 
Flame system: 
 Acetylene air burner100mmNitrous oxide acetylene flame head50mmIgnition dynamic baseline drift≤0.006A/30min(Cu) Characteristic viscosity≤0.025μg/ml/1%Related standard deviation of the accuracy≤0.5%(Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)Safety systemCan automatically cut off the gas when the pressure is not enough, the power is off, flam 
 
Graphite furnace:
 The highest temperature3000ºCThe largest temperature rising speed≥2000ºC/SCharacteristic quantityCd≤0.5×10-12g  Cu≤0.5×10-11gAccuracyCu≤3%  Cd≤3%Size and weight550mm×450mm×300mm 65KgSafety systemOver current protection
Low air pressure alarm / protection
Low cooling water flow alarm/protectionPower source and power220V±22V AC 7000W