Test Instrument

Lab Low Pressure High Alititude Instrument Test Chamber

Dongguan Xin Bao Instrument Co., Ltd.
  • Max Diameter of Workpiece:2000mm
  • Mass Scope of Workpiece:900kg
  • Type:Low Air Pressure Chamber Altitude Test Chamber
  • Maxcapacity:<1000KN
  • Accuracy Grade:0.5
  • Load Way:Electronic Load

Base Info

  • Model NO.:XB-OTS-252M
  • Loading Method:Dynamic Load
  • Display:Digital
  • Control:Computer Control
  • Weight:100-500Kg
  • Power Source:AC380V
  • Oil Cylinder Position:Under
  • Temp Range:-70c~+150c
  • Standard: :Ce, IEC, Mil
  • Product Name:Altitude Simulation Test Chamber
  • Inner Size (D*W*H)mm:800*700*900 mm
  • Temp Fluctuation:+,-0.5%
  • Cooling Rate:0.7c~1.0c,Min (Average)
  • Pressure Range:101kpa ~1kpa
  • Internal Chamber:Stainless Steel(SUS304)
  • Power:AC380V,50Hz, 3phase 4 Wires+ Ground Wire.
  • Customization Servise:Yes
  • Transport Package:Strong Wooden Case
  • Specification:(D*W*H)600*600*700mm
  • Trademark:OTS
  • Origin:China
  • HS Code:9031200000

Description

Specifications ModelXB-OTS-252MXB-OTS-504MXB-OTS-1000MXB-OTS-2160MInner Size (D*W*H)mm600*600*700800*700*9001000*1000*10001200*1400*1200Capacity(KW)11151618Chamber Volume(L)25250410002160 
 
 
 
 
SpecificationsTemp range-70ºC~+150ºC-70ºC~+150ºC-70ºC~+150ºC-70ºC~+150ºCTemp fluctuation±0.5%Temp deviation≥100ºC, ±3.0ºC(at 101kPa, ATM no load)<100ºC, ±2.0ºC(at 101kPa, ATM no load)Cooling rate0.7ºC~1.0ºC/min (Average)Pressure range101kPa ~1kPaPressure accuracy ±2kPa (101kPa ~ 40kPa),  ±0.5kPa (4kPa~40pKa ),   ±0.1kPa (4kPa~1kPa)Depressurization rate101kPa ~1kPa≤30mins (Room temp)Pressure recovery rate10kPa/Min(adjustable) 
 
MaterialExternal chamberCold-rolled steel sheet(Rust proof&Plastic Spray treated)/Stainless Steel plateInternal chamberStainless steel(SUS304)Pressurized partsHigh quality steel plate(Rust proof & Plastic spray treated)  Insulation materialHigh quality glass woolCooling
SystemCompressorSemi-Hermetic CompressorCooling wayWater cooledTemp regulating methodBalanced Temperature & Humidity Control System (BTHC)PumpRotary vane type vacuum pump HeaterIron-chrome wire heaterSight windowCircular viewing glass incorporating heat generatorTemp SensorPt-100ControllerTouch Screen Controller  Safety DevicesOver pressure, heat and current protection compressor, over temp. protection, over load protection for blower, hydraulic and water lack Protection, leakage protection, phase-lack protection, Pump failure alarmStandard AccessoryViewing window, Cable Port( 1 located on the left side), Baffle, Baffle frame*2  PowerAC380V/50Hz, 3Phase 4 wires+ Ground Wire. Note: 1. We reserve the right to change specifications without prior notice  2. Customized sizes and configurations are available
Standard configuration:
   Two sample holders; an observation window; φ25 or φ50 cable hole.
Optional parts:
   Communication interface RS232 or RS485, logger, micro printer, remote temperature and humidity control system (including printers, communications software, computer), mobile casters.
Structure:
Overall structure: the new design is easier to operate and maintain.
Water and electricity system is separated;
External humidifying design;
Three levels of overheat protection, easy access for refrigeration servicing, etc.
Material:
 
Tecumseh, Bock Compressor;
High quality electronic components (DANFOSS, ALCO, RANCO, SPORLAN, SAGINOMIYA, Schneider solenoid valve)
Standards Implemented and Met
B10592-89 Technical requirements for high and low temperature test chamber
GB2423.1-1989 Low temperature test
GB2423.2-1989 High temperature test mode

Thermostat regulator balanced manner BTPC thermostat (regulator) mode
Ambient temperature: 5 ºC ~ 35 ºC; Humidity: ≤90% RH; Atmospheric pressure: 86-106kPa
Material cold rolled steel shell, surface spray
SUS304 stainless steel inner tank
Rigid polyurethane foam insulation foam glass +
Regulator heater Nichrome heater
Fan shaft seal magnetic fluid flow fan
Refrigeration machinery contraction two binary cascade refrigeration (cooling)
Refrigeration compressor: Semi-hermetic compressors
Refrigerant:  R404a / R23
Controller Display: 7 "color screen, liquid crystal display (LCD)
Run the program mode, setting mode
Setting mode Touch (piezoelectric)
Program capacity 32 programs, each up to 800 steps (segments), maximum 999 cycles
Setting range operating temperature range adjustment device according to (limit 5 ºC, the lower limit of -5 ºC)
Display Resolution time: 0.1min; temperature: 0.1 ºC; pressure: 0.1kPa
Enter thermal resistance, pressure sensors
Anti-windup PID control mode
Power: AC380 (1 ± 10%) V (50 ± 0.5) Hz three-phase four-wire protective earth
Maximum power current: (A) 35 35

Safety devices
1. Power owed to reverse protection
2. No fuse switch
3. Dry compustion prevention switch
4. The water protection switch
5. Over-temperature protection
6. Motor overload protection
7. The high-pressure compressor protection
8. The overheating protection of compressor
9. The over-current protection of compressor
10. The over-current protection of vacuum pump
FEATURES
1. Temperature:  -70 ~170ºC
2. Altitude: ground to 100,000 feet
3.External pressure type box body structure and stainless steel tank design, rational air circulation system and scientific layout of heating, refrigerating system.
4.Adopts international brand of imported parts for cooling system to improve operation reliability
5. High precision and stable performance with touch screen controller, friendly Man-Machine interface.
6. Multiple layers of safe device , easy for maintenance.
7.Viewing window and cable ports are available.
8. Vacuum pump sized for climb/dive rate required by application

 Altitude chamber working principle

Packing & Shipping
1.Strong standard wood case, inner plastic films wrapped for Low Air Pressure Test Climatic Altitude Equipment
 
2. Usually we delivery the equipment by air,by sea or by express. it depends on customer 

About US

Our Services
We are here to help you. Services available:
Replacement manuals
Consumables and replacement parts
On-site maintenance and repair
Warranty support
Over-the-phone service support and advice
If you have an OTS chamber  , we will do our best to help you in whatever way we can.  
 Manuals: PDF files of manuals for current models are available free. Please contact us to request a copy. Older and printed copies at additional charge.
Service repair:  We prefer you contact us direct at the factory with any problems you may be experiencing.  We help diagnose the problem, arrange for any needed parts, and schedule a technician to do the repairs.  Rarely can a local contractor get all that done for you as quickly as we can.  We also guarantee the work 


RFQ
Q1 :  Can you customize the product?
   A:   Yes, we can customize any products according to requirements.
Q2 :  How to choose the most suitable products?
   A:   Our professional team will give you the best suggestion,just be  patiently to tell us the information about the parameters your required.
 Q3 :  Can you arrange installations and commissioning in our lab?
   A:   Firstly, for standard chamber we had 3-5 days commissioning before shipping , you just need link power in your lab, then can start to operate
     Secondly; we had detailed operation manual with chamber , any problems you can find and solve in it
      Thirdly, some of customized chambers we should to install and teach your team to operate .

If you want to know more information...
Please leave a message below and click " SEND "
Thanks for your time and attention~Thank you!