Laboratory ASTM D5453 Instrument for Total Sulfur in Diesel and Liquid Fuels
Chongqing Gold Mechanical & Electrical Equipment Co., Ltd.- Max Diameter of Workpiece:<500mm
- Mass Scope of Workpiece:<900kg
- Type:Universal Testing Machine
- Maxcapacity:<1000KN
- Accuracy Grade:0.5
- Control:Computer Control
Base Info
- Model NO.:GD-5453
- Weight:0-100Kg
- Power Source:AC220V
- Test Standard:UV Fluorescence ASTM D5453
- Total Sulfur in Diesel and Liquid Fuels:Ultra Low Sulphur Diesel Tester
- ASTM D5453 Sulfur Content:Trace-Level Total Sulfur Determination
- Measuring Range:0.1~10000mg,L
- Transport Package:Wooden Case
- Specification:Conforms to ASTM D5453
- Trademark:GOLD
- Origin:Made in China
- HS Code:9031809090
- Production Capacity:100sets,Month
Description
Basic Info.
Model NO. GD-5453 Weight 0-100Kg Power Source AC220V Test Standard UV Fluorescence ASTM D5453 Total Sulfur in Diesel and Liquid Fuels Ultra Low Sulphur Diesel Tester ASTM D5453 Sulfur Content Trace-Level Total Sulfur Determination Measuring Range 0.1~10000mg/L Transport Package Wooden Case Specification Conforms to ASTM D5453 Trademark GOLD Origin Made in China HS Code 9031809090 Production Capacity 100sets/MonthProduct Description
Laboratory astm d5453 instrument for total sulfur in diesel and liquid fuelsHigh Sensitivity: The astm d5453 instrument is used to determine the total sulfur content by ultraviolet fluorescence method. It improves the ability of anti-jamming, and avoids the complicated operation of titration pool and factors of instability which used Coulometry. So the sensitivity of the instrument is greatly improved.
The key components of the system adopt imported components to protect the detector light leakage and steam leakage and makes the machine has a reliable guarantee.
Standards:
SH/T 0689-2000, ASTM D5453-2006 , EN ISO 20846:2004, GB 17930-2011, GB 19147-2009, ISO 20846-2011, BS EN 15486-2007
System Configuration:
Standard configuration: Printer+Computer+GD-5453+Liquid Injector
Other optional parts: Solid sample injector, Gas sample injector
Easy to Operate:
Based on Windows (7, XP, Me, 98 ) English user interface to make the operation more convenient and fast. The operator only needs to click the mouse, and it can complete all of the parameter settings. The data collecting, processing, storage and printing are fully controlled by computer.
Technical Parameters:
Sample types: Solid, Gas and Liquid
Method of determination: Ultraviolet fluorescence method (S)
Sample injection quantity: Solid: 1-20mg; Liquid: 5-50μL; Gas: 1-5mL
Measuring range: 0.1~10000mg/L(High concentration should be diluted, Low concentration gas sample is up to 0.1ppm)
Measuring accuracy:
Concentration values (ppm) | Injection quantity (µL) | RSD (%) |
0.2 | 20 | 50 |
5 | 10 | 10 |
50 | 10 | 5 |
100 | 10 | 3 |
5000 | 10 | 3 |
Temperature range: room temp~1150C
Temperature-controlled precision: ±1C
Air supply requirements: High-purity argon: above 99.9%
High purity oxygen: above 99.9%
Power supply: AC220V±22V, 50Hz±0.5Hz, 1500 W
Dimension: Host: 305(W)×460(D)×440(H)mm
Temp controller: 550(W)×460(D)×440(H)mm
Weight: Host: 20kg; Temp controller: 40kg