Test Instrument

Automatic Pensky-Martens Closed-Cup Tester for Residual Fuel Oils, Cutback Residua, Used Lubricating Oils

Chongqing Gold Mechanical & Electrical Equipment Co., Ltd.
  • Test Method:Pensky-Martens Closed Cup Method ASTM D93
  • Testee:Biodiesel, Distillate Fuels, Used Lube Oil, etc.
  • Flash Point Determination:Ambient to 300c
  • Igniting Mode:Electric Ignition
  • Warranty:1 Year
  • Transport Package:Plywood

Base Info

  • Model NO.:GD-261A
  • Origin:Made in China
  • HS Code:9031809090
  • Production Capacity:100sets,Month

Description

Basic Info.

Model NO. GD-261A Origin Made in China HS Code 9031809090 Production Capacity 100sets/Month

Product Description

GD-261A Automatic Pensky-Martens Closed-cup Tester for residual fuel oils, cutback residua, used lubricating oils



Introduction:

Flash point is an attribute associated with events far more ferocious and destructive than a simple flash of light. Defined as the lowest temperature of a liquid at which its vapors will form a combustible mixture with air, flash point is a convenient and reliable classification of flammability for numerous petroleum liquids and related substances. 

Pensky-Martens test procedures can  determine the flash point of petroleum products in the 40°C to 370°C range by either manual or automated closed-cup apparatus, as well as determining the flash point of biodiesel in the 60°C to 190°C range by an automated closed-cup apparatus.

The PMCC Flash Point Tester is used to test the biodiesel, distillate fuels, new lubricating oils, residual fuel oils, cutback residua, used lubricating oils, mixtures of petroleum liquids with solids, petroleum liquids that tend to form a surface film during testing.


Features:

1. Simulation tracking displays function curves of heating up and test time.
2. Equipped with the English prompt function of mis-operation.
3. Equipped with prompt function of test date,test time and other parameters.
4. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.
5. Differential coefficient detection.Automatically correct the system deviation.
6. Automatically open the lid,detect the flash point and print test data.
7. The test arm rises up and lowers down automatically.
8. Igniting with electric igniting gun,forced air cooling.


Speficiations:

1. Power supply: AC(220±10%)V,  50Hz.
2. Flash point determination: Range: Ambient to 300ºC
                                                Repeatability: ≤ 2ºC
                                                Reproducibility: ≤ 4ºC
                                                Accuracy: 0.1ºC
3. Heating rate: Procedure A: (5~6)ºC/min
                            Procedure B:(1~1.6)ºC/min
                            Automatic control and manually adjustable.
4. Stirring rate: Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
                          Automatic control and manually adjustable.
5. Igniting mode: (1) Igniting source: gas(or other civilian fuels,the same below)
                             (2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
6. Ambient temperature: ≤ 40ºC
7. Relative humidity: ≤ 80%
8. Maximum power consumption: 500W



Welcome to visit us!

Automatic Pensky-Martens Closed-Cup Tester for Residual Fuel Oils, Cutback Residua, Used Lubricating Oils