Test Instrument

Petroleum Testing Instruments Manually Cleveland Open Cup Flash Fire Point Tester for Sale

Chongqing Gold Mechanical & Electrical Equipment Co., Ltd.
  • Max Diameter of Workpiece:<500mm
  • Mass Scope of Workpiece:<900kg
  • Type:Cleveland Open Cup Flash Point Test
  • Maxcapacity:<1000KN
  • Accuracy Grade:0.5
  • Load Way:Electronic Load

Base Info

  • Model NO.:GD-3536
  • Loading Method:Dynamic Load
  • Display:Digital
  • Control:Computer Control
  • Weight:0-100Kg
  • Power Source:AC220V
  • Oil Cylinder Position:Under
  • Power Supply:AC(220±10%)V, 50Hz.
  • Heating Device:Quartz Tube Furnace Heating, No Naked Fire
  • Test Flame Applicator:It Applies The Test Flame Automatically.
  • Thermometer:(-6~400)&Ordm
  • Igniting Device:1.Ignition Source: Coal Gas 2.Lame Diameter Is 3.2
  • Ambient Temperature:(-10-50)ºC
  • Relative Humidity:≤85%
  • Maximum Power Consumption:650W
  • Transport Package:Wooden Case
  • Specification:350mm*290mm*350mm
  • Trademark:GOLD
  • Origin:China
  • Production Capacity:200units,Month

Description

Basic Info.

Model NO. GD-3536 Loading Method Dynamic Load Display Digital Control Computer Control Weight 0-100Kg Power Source AC220V Oil Cylinder Position Under Power Supply AC(220±10%)V, 50Hz. Heating Device Quartz Tube Furnace Heating, No Naked Fire Test Flame Applicator It Applies The Test Flame Automatically. Thermometer (-6~400)&Ordm Igniting Device 1.Ignition Source: Coal Gas 2.Lame Diameter Is 3.2 Ambient Temperature (-10-50)ºC Relative Humidity ≤85% Maximum Power Consumption 650W Transport Package Wooden Case Specification 350mm*290mm*350mm Trademark GOLD Origin China Production Capacity 200units/Month

Product Description

Petroleum Testing Instruments Manually Cleveland Open Cup Flash Fire Point Tester for Sale



Introduction:

This cleveland open cup flash and fire point tester is designed and made as per standards GB/T3536-2008 Petroleum Products -  Determination of Flash and Fire Points - Cleveland Open Cup Method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79ºC and below 400ºC except fuel oils.

 

Features:

I. Main technical features

1. It adopts special heating furnace to ensure the safety of test. The heating power is continuously adjustable.It can meet requirements of test.

2. The cleveland open cup flash and fire point tester adopts desktop structure.The stainless steel working 
table and furnace shell make it beautiful and elegant.

3. The instrument is fully self-contained complete.The operator can do determination as long as connecting with coal gas or other civil gas.

4. It equipped with flame extinguishing cover which accord with requirements of test

 

II. Main technical specifications

1. Power supply:  AC(220±10%)V, 50Hz.

2. Heating device: Quartz tube furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W to 600W.

3. Test flame applicator: It applies the test flame automatically.

4. Thermometer: (-6~400)ºC. It is the same as thermometer ASTM 11C.

5. Igniting device: (1) Ignition source: coal gas(or civil gas)

               (2) Flame diameter is 3.2mm~4.8mm

6. Ambient temperature: (-10∼50)ºC

7. Relative humidity: ≤85%

8. Maximum power consumption:  650W

9. Dimension: 350mm×290mm×350mm (thermometer is not included)


III. Detail Images:
Petroleum Testing Instruments Manually Cleveland Open Cup Flash Fire Point Tester for Sale

Petroleum Testing Instruments Manually Cleveland Open Cup Flash Fire Point Tester for Sale

Other Related Products: