Test Instrument

Benchtop Xrf X-ray Fluorescence Sulfur Analysis ASTM D4294 Analyser for Bunker Fuel Oil

Chongqing Gold Mechanical & Electrical Equipment Co., Ltd.
  • Test Method:ASTM D4294 Xrf
  • Transport Package:Wooden Case
  • Origin:China

Base Info

  • Model NO.:GD-4294
  • HS Code:9031809090
  • Production Capacity:200PCS,Month

Description

Basic Info.

Model NO. GD-4294 HS Code 9031809090 Production Capacity 200PCS/Month

Product Description

GD-4294 BENCHTOP XRF X-ray Fluorescence Sulfur Analysis ASTM D4294 Analyser for Bunker Fuel Oil


Features and Advantages of X-ray Fluorescence Sulfur Analysis ASTM D4294 Analyser for Bunker Fuel Oil

1. The XRF ASTM D4294 Sulfur Analyser adopts electromechanical integrated micro-computer design, 8-inch wide viewing angle capacitive touch screen (1024*768) display, backlight brightness can be set by the software, no keyboard is needed, and the interface is simple and beautiful.
2. Wide variety of detection, wide measuring range, fast analysis speed, less sample consumption.
3. Using the fluorescence intensity ratio analysis method, the temperature and air pressure are automatically corrected, and the hydrocarbon ratio (C/H) can also be corrected.
4. The automatic diagnosis function of the instrument determines the working state and electrical parameters of the instrument.
5. Using the disposable Mylar membrane sample cup, it can avoid cross-contamination; the sample cup is made of multi-function pressing, which is quick and convenient. 6. The sample stage is positioned accurately, it is very convenient to place the sample and replace the oil leakage prevention component, and avoid the possibility that the detection system is contaminated.
7. The instrument has a large amount of data storage. The default storage is 4096 content analysis results, 8192 count measurement data and 16 instruments calibration result data, the data can be queried, and it can also be uploaded to the computer through the RS-232 standard serial communication port.
8. The instrument has the function of automatic peak stabilization, when the performance of the detector decreases, the system can automatically adjust the high voltage and correct the error.
9. The Bunker Fuel Oil Sulfur Analyser automatically selects the working curve by default, without user intervention.
10. The user can select the measurement result unit, ppm or (m/m)%.
11. Safety protection measures of X rays can prevent ionizing radiation damage to individuals or groups.


Specifications:
(1) Measurement range: 50 ppm to 5%.
(2) Repeatability (r): <0.02894(S+0.1691).
(3) Reproducibility (R): <0.1215(S+0.05555).
(4) Detection limit: 10 ppm.
(5) Oil sample quantity: 3 ml (it is equal to sample depth of 3 mm~4 mm).
(6) Measurement time: it can be set to 30s or 60s or 90s or 120s or 150s.
(7) It can make determination automatically for single sample. Replication times can set 1,2, 3, 5, 10 at random; it will show average value and standard deviation at end of measurement.
(8) Calibration curve numbers: it can save 10 calibration curves.
(9) Working condition:
      a. Ambient temperature: 5~30ºC.
      b. Relative humidity: <=85% (30 ºC).
(10) Power supply: AC 220V±20V, 50 Hz/60Hz; Rated power: 50 W.
(11) Size and weight: 460mm×260mm×260mm; 12 kg.


Welcome to visit us!